Technical characteristics and internals of NanoVNA.

NanoVNA Technical Specifications and Hardware Architecture

The hardware of the NanoVNA represents a masterpiece of cost optimization. The developers managed to achieve excellent radio frequency (RF) parameters based on accessible and inexpensive general-purpose radio components.

Technical Specifications of the Classic NanoVNA

Device Parameter Technical Specifications of NanoVNA V1
Frequency Range 50 kHz – 300 MHz (base), up to 900 MHz (harmonic mode), up to 1500 MHz (extended)
Number of Scan Points 101 points (fixed hardware limitation in internal memory)
RF Output (Port CH0) -13 dBm nominal (maximum up to -9 dBm) into a 50 Ohm load
S21 Dynamic Range 70 dB (50 kHz – 300 MHz), 50 dB (300 – 600 MHz), 40 dB (600 – 900 MHz)
Measurement Port SWR < 1.1 across the entire working frequency range
Frequency Error < 2.5 ppm (reference oscillator frequency stability < 0.5 ppm)
Device Display 2.8 inches, TFT, resistive touchscreen, 320 x 240 pixel resolution
Communication & Power Interface USB Type-C, USB CDC class (virtual COM port)
Built-in Battery Lithium-polymer, 3.7 V, capacity ranging from 400 mAh to 450 mAh

Hardware Architecture and Circuit Solutions

At the core of the classic NanoVNA lies a superheterodyne architecture featuring a single switched receiver. The main functional units of the device are implemented using the following components:

  1. Signal Synthesizer (Si5351A): The integrated dual-channel Si5351A clock generator chip produces the high-frequency signals for the reference Local Oscillator (LO) and the test stimulus. The signal from the Si5351 is a square wave, which inherently contains a large number of odd harmonics. At frequencies above 300 MHz, the NanoVNA utilizes the 3rd and 5th harmonics of the synthesizer to scan the RF range. This ingenious approach allowed expanding the measurement limit up to 900–1500 MHz without increasing the circuit cost.
  2. Directional Coupler and Mixers: An RF bridge (Wheatstone bridge) extracts the reflected signal. The coupled signal is fed into AD8342 or classic SA602 mixers, which downconvert it to a low Intermediate Frequency (IF) of around 5–12 kHz.
  3. Measurement Path and ADC (TLV320AIC3204): A high-quality, low-power stereo audio codec (TLV320AIC3204) serves as the Analog-to-Digital Converter (ADC) for the IF signals. It digitizes the low-frequency IF signals at a 48 kHz sampling rate, transmitting the data via the I2S bus directly to the controlling microcontroller.
  4. Microcontroller (STM32F072C8T6 or GD32): This chip acts as the brain of the device. It performs overall device control, processes the digitized quadrature I/Q vectors, calculates the complex S-parameters, and renders the graphs onto the ILI9431 LCD screen. Since the signal processing is executed entirely in the digital domain, it ensures high parameter stability.


Lineup of device modifications: NanoVNA-H, NanoVNA-H4, and NanoVNA V2 SAA-2

Comments

Popular posts

Antenna calculation

Formulas for Dipole and Vertical Antennas Half-wave antenna in free space (15006 / F) cm Practical half-wave antenna (up to 30 MHz) (14274 / F) cm Half-wave antenna (50 to 144 MHz) (14030 / F) cm Half-wave antenna (above 144 MHz) (14233 / F) cm Quarter-wave antenna (7137 / F) cm Full-wave antenna length (30653 / F) cm Enter frequency (MHz): Calculate ...

NanoVNA. High-precision calibration.

High-Precision SOLT (OSLT) Calibration A Vector Network Analyzer (VNA) is a precision instrument whose readings directly depend on calibration. Without calibration, the VNA displays not the parameters of the measured antenna or filter, but the characteristics of the entire measurement line (including the parasitic capacitance and inductance of the connecting cables and connectors). The OSLT Calibration Procedure To perform the calibration, a set of calibration standards from the kit is used: OPEN (open connector, infinite impedance), SHORT (short-circuited connector, 0 Ohms), and LOAD (matched 50-ohm non-inductive load). It is vital to perform the calibration exactly at the end of the cable to which the Device Under Test ( DUT ) will be directly connected. Crucial Interpolation Rule: When changing the frequency range ( START/STOP ), the device automatically interpolates the calibration matrix, and the calibration status is disp...

Dipole antenna simulator.

Dipole Antenna Simulator Interactive E-Field, Current & Impedance Visualizer How to use this simulator: Custom Input: Type your exact Operating Frequency or Physical Length into the input boxes, or use the sliders to experiment dynamically. Visualizing Standing Waves: The wire in the diagram takes up the full width for clarity. The RED graph shows Current (I) magnitude, and the BLUE graph shows Voltage (V) magnitude along the wire. Notice how Current is always zero at the antenna tips! Experiment: Increase the length beyond 1 wavelength (1.0 λ) and watch how the radiation pattern splits into multiple lobes. Auto Tune: Click the green button to instantly calculate the perfect resonant half-wave length (0.48 λ) for your chosen frequency. ...